Scope of the accreditation

Calibration laboratory for Meteorological Instruments is accreditated with according ISO 17025 since 2002.

item

Type of the instrument Range of measurement Expanded ucertainry U
(k=2)
established methods Others specifications  
method document  
 
1 in-glass thermometers (-30 až +40) °C 0,07 °C immediate comparison with standard thermometers MK 012/97 medium is water and ethanol  
2 resistance thermometers (-30 až +40) °C 0,05 °C

immediate comparison with temperature standard: SPRT Rosemount and multimeter  Hart Scientific

MK 013/97 medium is water and ethanol  
3 thermographs (-30 až +40) °C 0,3 °C immediate comparison with temperature standard: PRT and datalogger MK 004/96 in climat chamber, medium is air  
4 mechanical hygrometers (10 - 95)% 5,0 % immediate comparison with humidity standard MK 001/96 in climat chamber  
5 relative humidity sensors (10 - 95)% 2,5 % immediate comparison with humidity standard MK 001/96 in climat chamber  
6 relative humidity sensors (5 - 95)% 2,5 % immediate comparison with humidity standard: General Eastern D2 MK 005/01 in generator General Eastern C1  
7 hygrographs (10 - 95)% 3,0 % immediate comparison with humidity standard MK 001/96 in climat chamber  
8 air pressure sensors (650 - 1050) hPa 0,1 hPa immediate comparison with pressure standard- kalibrátor DHI PPC2+ MK 023/98    
9 aneroids (650 - 1050) hPa 0,01 % of the range immediate comparison with pressure standard- kalibrátor DHI PPC2+ MK 023/98    
10 anemometers (0,3 - 50 ) m.s-1 (0,017v + 0,1) m.s-1

immediate comparison with standards: thermal, vane probe and Pitot tube

MK 031/99 in wind tunnel  
11 tipping bucket rai gauges (5 - 500)ml 2 % immediate comparison with volume standard at several rain intensities  MK 043/04 rain intensity generator  
12 weight rain gauges 1 g - 7 kg 0,5 g immediate comparison with weight standards MK 044/04 differential method  
13

Analysers of

NO-NO2-NOx

(0,5 - 1000) nmol.mol-1 (0,04 hm+3) nmol.mol-1 immediate comparison with standard analyser MK 052/03 with calibration gas mixture  
14

Analysers of

SO2

(0,5 - 1000) nmol.mol-1 (0,05 hm+3) nmol.mol-1 immediate comparison with standard analyser MK 051/03

with calibration gas mixture

 
15

Analysers of

O3

(0,5 - 500) nmol.mol-1 (0,03 hm+3) nmol.mol-1 immediate comparison with standard analyser MK 053/03 generator of the  O3  
16

Analysers of

CO

(0,04 - 50) µmol.mol-1 (0,02 hm+1) µmol.mol-1 immediate comparison with standard analyser MK 054/03

with calibration gas mixture

 

Contact:

Jan.Danc@shmu.sk, Jozef.Lengyel@shmu.sk